ANY PATIENT, ANY SIDE, IT FITS!
Video INSTRUCTIONS
APII Prep with Dr. Leah Walker
APII Bonding with Dr. Leah Walker
Butter Base with Dr. Leah Walker
Debond with Dr. Leah Walker
APII Assembly
APII Debond
INSTRUCTIONS
SIZE - CUT - CRIMP - BOND
SIZE
SIZE - CUT - CRIMP - BOND
Use the Sizing Template and align the “M/B Groove” line on the buccal groove of the maxillary 1st molar. Then measure the height of the contour on the cuspid or bicuspid to determine the length of the Maxillary Arm.
Align the permanent marking on the cuspid/bicuspid hook with the required maxillary arm length. The “CUT-LINE” locates where to cut the maxillary arm.
CUT
SIZE - CUT - CRIMP - BOND
A Hard Wire Cutter or Lab Cutter is recommended for cutting the Maxillary Arm to length. The permanent markings on the Maxillary Arm are 2 mm apart. A 1 mm increment in length reduction is accomplished by cutting the Maxillary Arm at the mid-point between two permanent markings.
CRIMP
SIZE - CUT - CRIMP - BOND
Ensure the Maxillary Arm is fully seated in the Receiver.
Use an Orthodontic Hard Wire Cutter or Lab Cutter to crimp the Receiver on its O-G sides at two locations:
-
Near the funnel entry to the Receiver
-
At the mid-point of the Receiver
Bond
SIZE - CUT - CRIMP - BOND
The APII device is ready to be bonded. Factory-installed Ligature Ties are attached to the Molar Attachment distal tie wings to avoid accidental disengagement of the Maxillary Arm from the Molar Attachment during bonding. Make sure to remove them after bonding is complete.
Download APII Instruction Flyer Here: